MedWOW / Медоборудование / Технологии визуализации / МРТ / Siemens / MAGNETOM Open Viva / Технические характеристики производителя
 

Технические характеристики производителя - MAGNETOM Open Viva, Siemens

Обращаем Ваше внимание: None of the equipment posted for sale on MedWOW.com is owned by MedWOW, should you have any questions regarding a specific item, please direct them to the appropriate seller by making use of the available communication channels on the items page.
MODEL

MAGNETOM Open Viva

MAGNETResistiveC-shaped
POWER NEEDED
Line voltage, V (POWER NEEDED)380/400/420/440/480
Kva (POWER NEEDED)35
CLINICAL USE (POWER NEEDED)Whole body
CARDIAC GATING
Ecg/peripheral (CARDIAC GATING)Yes/yes
IMAGING MODES
Single (EYEPIECE)Yes
Multislice (IMAGING MODES)Yes
Volume study (IMAGING MODES)Yes
SURFACE COILS
Spine (SURFACE COILS)CP flexible spine
Knee (SURFACE COILS)21cm multipurpose
Neck (SURFACE COILS)16/21 cm multipurp
Tmj (SYSTEM CAPABILITIES)
Extremity (SURFACE COILS)16/21 cm multipurp
Head (SURFACE COILS)CP head
Shoulder (SURFACE COILS)16/21 cm multipurp
Additional (Alarms, high/low)45cm multipurpose, 35cm multipurpose for MR-guided procedures
BORE DIAMETER or W x H, cm (SURFACE COILS)MRT (Open Magnet)
DICOM 3.0 COMPATIBLE (POWER NEEDED, VAC)Optional
GRADIENT SUBSYSTEM
Strength, mt/m (GRADIENT SUBSYSTEM)15
Rise time to 10mT/m, msec (GRADIENT SUBSYSTEM)0.6
Cpu (NETWORK/DATABASE)Turbo SPARC-based host
Memory size, mb (GRADIENT SUBSYSTEM)64
ARRAY PROCESSOR (EXTERNAL BEAM)SMI-5 (Siemens Medical Imager)
Memory size, mb (GRADIENT SUBSYSTEM)146
Storage media/size (GRADIENT SUBSYSTEM)1.7GB MOD drive
Image storage capacity (GRADIENT SUBSYSTEM)~11,000 (256 x 256)
ELECTROMAGNETIC FIELD STRENGTH, T (5-GAUSS FRINGE FIELD)0.2
OPTIONSPACSnet Ethernet, MR-guided procedures kit, comfort kit, cardiac evualuation package, extra memory, hard disk expansion, Prominence work- station
SHIMMING (CRYOGEN USE, L/hr)Passive, active
Diameter, cm (VACUUM)
Body coil (DIAMETER, cm)41pole to pole
Head coil (DIAMETER, cm)20.3x 17
Reconstruction time
Single slice, sec (RECONSTRUCTION TIME)<0.65
Multislice, sec (RECONSTRUCTION TIME)<0.65/slice
Volume, sec (RECONSTRUCTION TIME)<0.65/slice
Cooling method (X-RAY TUBE)Closed loop, chilled water
Chilled h2o, l/min (POWER NEEDED)17-40, varies w/temp
MAGNET WEIGHT, kg (CRYOGEN USE, L/hr)12100
DIMENSIONS (HXWXD),CM (CRYOGEN USE, L/hr)195x 120 x 200
AngiographyOptional MT, TONE, 2-D & 3-D TOF; Turbo MRA, PC (WIP)
IMAGING
Pulse sequences (IMAGING)SE, IR, True IR, Turbo IR, GE, 2-D & 3-D FLASH, 2-D & 3-D FISP, 3-D PSIF, 3-D DESS, 2-D True FISP, TSE, Turbo FLASH, combination echo
Repetition time, msec (IMAGING)28minimum SE, 11minimum GE
Echo time, msec (IMAGING)15mininum SE, 4minimum GE
Inversion time, msec (IMAGING)35minimum
Slice thickness, mm (GANTRY)2-60 (2-D), 0.5-50 (3-D)
Fov, cm (IMAGING)4-40
Fov offsets (IMAGING)±20
Scan orientations (IMAGING)Axial, sagittal, and coronal orthogonal; oblique; double oblique; MAO
Measuring matrix (IMAGING)Up to N x 1024, N = 32-512 in increments of 1
Display matrix (IMAGE DISPLAY)1024x 1024
Pixel intensity (IMAGING)1-4,096
Spatial resolution, mm (IMAGING)0.31,in plane
BORE FEATURES (DIAMETER, cm)Open 270 deg patient access, removable table, intercom, nurse-call bulb, opt music headphones
CONFIGURATIONOpen
OTHER ATTRIBUTES (Interference compensation)Integrated MIP and MPR; standard integrated image noise filters, VCR interface, cine display, remote diagnostics package, quantitative analy- sis; 2 coils plug in simultaneously.
FDA CLEARANCE (Interference compensation)Yes
CE MARK (MDD) (Interference compensation)Yes
MARKETING REGION (Interference compensation)Worldwide
Вы также можете быть заинтересованы в следующих технических характеристик производителя:
Ограничение ответственности

Информация на этой странице может быть неполной, содержать ошибки или неточности.


МРТ, Siemens, MAGNETOM Open Viva
View More Items