MODEL

HiSpeed CT/i

X-RAY TUBE
X-RAY TUBE ANODE
Heat storage, hu (X-ray tube anode)3500000
Heat dissipation rate, hu/min (X-RAY TUBE)820000
Tube cooling (X-ray tube anode)Oil/air
Tube focal spot, mm (X-ray tube anode)0.4x 0.9, 0.6 x 0.9
Optional x-ray tubes (X-ray tube anode)Performix 630; 6,300,000HU
POWER NEEDED480VAC, 50/60 Hz, 3-phase delta
Type (CONFIGURATION)Single slice helical
# of slices (X-ray tube anode)1
DICOM 3.0 COMPATIBLE (POWER NEEDED, VAC)Yes
HELICAL SCANNING (DISPLAY)Yes
Max scan time, sec (DISPLAY)60
Max scan volume, cm (DISPLAY)170
Spatial resolution, lp/cm (DISPLAY)Same as axial
Pitch (DISPLAY)0.4:1 to 3:1
Reconstruction time per image, sec (DISPLAY)3 (1.75optional)
Output, kw (X-RAY GENERATORS)48,optional 53
Kvp range (GENERATOR)80.100.120.140
Ma range (IMAGING SYSTEM)10440 with Performix
3-D PACKAGE (DISPLAY)Optional
FLOOR SPACE NEEDED
Size, m2 (FLOOR SPACE NEEDED)28 (45optimal)
Image storage
Hard disk, MB4.3GB
No. online images (IMAGE STORAGE)3,700 (512x 512), 450raw data
Archive (IMAGE STORAGE)2.3GB MOD, DICOM 3.0
DISPLAY
MONITOR20"(2)
Matrixes, pixels (DISPLAY)1024x 1024
Range of ct numbers (DISPLAY)-1,024to +3,071
Image enlarging scale (DISPLAY)Up to 8x
Max # slices displayed simultaneously (DISPLAY)16
GENERATOR
CT SIMULATION (DISPLAY)Optional
IMAGE PROCESSING (MAXIMUM DISPLAY)
Cpu (NETWORK/DATABASE)Silicon GraphicsMIPS 4400/250 MHz
Scan fovs, cm (IMAGE PROCESSING)25.48
Image reconstruction matrixes (IMAGE PROCESSING)256x 256, 512x 512
Reconstruction time
Per slice, sec (Reconstruction time)3,optional 1.8
For localization scan, sec (Reconstruction time)Real time
TABLE FEATURES
Range of movement
Vertical, cm (Range of movement)51-107
Longitudinal, cm (Range of movement)170
Scanning range,cm (Range of movement)170
Max load capacity with accuracy, kg (Range of movement)180 (±0.25 mm), 205 (±1 mm)
PERFORMANCE
Minimum interscan time, sec (PERFORMANCE)1
Dynamic scan pace (PERFORMANCE)30scans/min
High-contrast spatial resolution 0%MTF, lp/cm (PERFORMANCE)15
50%MTF, lp/cm (PERFORMANCE)8.5
Low-contrast resolution, mm at %at £4 rads (PERFORMANCE)2.5mm at 0.25% at 4rads AAPM w/ATS
Noise, %at £2.5 rads (PERFORMANCE)0.33@ 2.5 rads
CT ANGIOGRAPHY (DISPLAY)Part of 3-D option
GANTRY
Geometry (GANTRY)Rotate-rotate, low-voltage slip ring
Number and type of detectors (GANTRY)840+ 12 reference, HiLight ceramic
# detection channels (GANTRY)744
Rotation times, sec 360 (GANTRY)1.2.3.4; 0.8optional
Partial (GANTRY)0.5
Slice thickness, mm (GANTRY)1,3,5,7,10
X-ray fan beam angle, ° (GANTRY)49
Gantry angle deg (GANTRY)-30to +30
Gantry size, hxwxd,cm (GANTRY)208x 223 x 109
Gantry weight, kg (GANTRY)2979
Gantry opening, cm (GANTRY)70
Scan localizer (GANTRY)Laser
OTHER ATTRIBUTES (Interference compensation)Autoscan, autovoice, autofilm, autostore, autotransfer; Smart- Prep, ProtocolPro, view/edit wizard, DynaPlan Plus ImageWorks, ProView, PMR ConnectPro, Sherlock CBT, bar- code reader, Smart- Score, SmartView, Fast Recon.
FDA CLEARANCE (Interference compensation)Yes
CE MARK (MDD) (Interference compensation)Yes
MARKETING REGION (Interference compensation)Worldwide
Вы также можете быть заинтересованы в следующих технических характеристик производителя:
Ограничение ответственности

Информация на этой странице может быть неполной, содержать ошибки или неточности.


Сканер КТ, GE Healthcare, HiSpeed CT/i
View More Items